Philips Medical Systems (Cleveland) Inc: Medical Device Recall in 2013 - (Recall #: Z-0335-2014)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

Brilliance iCT Computed Tomography X-Ray System, Product Usage: The Brilliance iCT is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.

Product Classification:

Class II

Date Initiated: October 11, 2013
Date Posted: November 27, 2013
Recall Number: Z-0335-2014
Event ID: 66601
Reason for Recall:

If the couch, with the footrest extension attached, is positioned where the table travel is within the bore of the gantry, and the user begins to lower the couch, the footrest extension can contact the gantry cover or come in contact with the user due to the length of the accessory.

Status: Terminated
Product Quantity: 9 units
Code Information:

Model #728306, Serial #'s: 100016, 100018, 100019, 100020, 100021, 100022, 100023, 100024 and 100025.

Distribution Pattern:

Worldwide distribution: USA (nationwide) Distribution in the states of DC, GA, IL, IN, NY and OH and the countries: Germany, Israel and Netherlands.

Voluntary or Mandated:

Voluntary: Firm initiated