Philips Medical Systems (Cleveland) Inc: Medical Device Recall in 2013 - (Recall #: Z-0950-2013)
See the recall detail below. You can also see other recalls from the same firm in 2013.
Brilliance 64 and Ingenuity Computed Tomography X-Ray Systems The Brilliance 64 and Ingenuity CT are whole body Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories.
Class II
This field change order is being released to update software and customer release notes to the affected installed base.
Brilliance 64: 728231 & Ingenuity CT: 728326, 728323. System Code # 728231, Serial #: 9654, 91003, 95130, 95157, 95504 System Code # 728326, Serial #: 32000,300003,300004,300005,300010,300011,300012,300014,300015,300018,300027,300041,300053,300059,300066,300070,300071,300073,300074,300078,300079,300080,300082,300084,300086,300091,300095,300101,300102,300112,300127,300128,300129,300131,300132,310064 System Code # 728323, Serial #: 320033
Worldwide Distribution - USA including AR, AZ, IN, KS, LA, MA, MN, MS, MT, NJ, NY, NY, OH, OR, PA, TN, TX, VT, WV, WY, and worldwide including the contries of Australia, Belgium, China, France, Germany, Israel, Japan, Norway, Sweeden, and Switzerland.
Voluntary: Firm initiated