Philips Medical Systems (Cleveland) Inc: Medical Device Recall in 2016 - (Recall #: Z-1504-2016)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2016.

Data Source: FDA.
Product Description:

Extended Brilliance Workspace NM Nuclear medicine image display and processing application suite. AutoSPECT Pro Software Application. Philips Medical Systems, Cleveland, OH. Provides software applications used to process, analyze, and display medical images/data.

Product Classification:

Class II

Date Initiated: March 4, 2016
Date Posted: April 27, 2016
Recall Number: Z-1504-2016
Event ID: 73776
Reason for Recall:

The AutoSPECT Pro application was only designed to reconstruct cardiac SPECT data obtained with detectors positioned at 90¿ or 180¿ relative to one another. However, certain gamma cameras allow for other relative detector angles. Data acquired at these other angles will not be correctly reconstructed by AutoSPECT Pro and the results will likely be erroneous.

Status: Terminated
Product Quantity: 126 Units
Code Information:

M/N 882489; S/N: 66794, 66798, 66836, 66009,66011, 67154, 66828, 67058, 67050, 67031, 67284, 66332, 66335, 66397, 67026, 67040, 66084, 66089, 66092, 67007, 67011, 67226, 66956, 66454, 66447, 66413, 66129, 66128, 66300, 66023, 11728, 66000, 66938, 67270, 67252, 66571, 66443, 66324, 66869, 68477, 66682, 66815, 66284, 67263, 66265, 66277, 66880, 66892, 66258, 66079, 66181, 66451, 66627, 66187, 66806, 67016, 67008, 66706, 66745, 66817, 66822, 66910, 66908, 66811, 67087, 67094, 66241, 66825, 66632, 66624, 66854, 66689, 66690, 66906, 66792, 67173, 67069, 67166, 68290, 67182, 66785, 67292, 67286, 66768, 66769, 66550, 66113, 67220, 67193, 67186, 66657, 66668, 66564, 66065, 67065, 67208, 67204, 66818, 66819, 67122, 66818, 66819, 66456, 67212, 66019, 67057, 67172, 67157, 67167, 67162, 66737, 68361, 66183, 66368, 66387, 66376, 66415, 66136, 66137, 66960, 66958, 67142 & 67165.

Distribution Pattern:

Worldwide Distribution -- US, including the states of CO, DE, FL, GA, IL, IN, KS, MO, NC, NV, NY, OH, OR, PA, TN & TX; and, the countries of Australia, Austria, Belgium, Canada, China, Denmark, Egypt, France, Germany, Greece, Ireland, Israel, Italy, Lebanon, Malaysia, Saudi Arabia, Slovenia, South Africa, Switzerland, Turkey, & United Kingdom.

Voluntary or Mandated:

Voluntary: Firm initiated