Philips Medical Systems (Cleveland) Inc: Medical Device Recall in 2016 - (Recall #: Z-2109-2016)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2016.

Data Source: FDA.
Product Description:

Big Bore Oncology tomography X-ray system scanner, model number 728243. The Big Bore Oncology scanner is a whole body computed tomography X-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes

Product Classification:

Class II

Date Initiated: June 14, 2016
Date Posted: July 6, 2016
Recall Number: Z-2109-2016
Event ID: 74507
Reason for Recall:

Socket Head Cap Screws (SHCS) used in the Z-Axis Adjustment Plate were found to be made of stainless steel rather than the specified alloy steel.

Status: Terminated
Product Quantity: 47 units
Code Information:

Big Bore Oncology System Code 728243; Serial numbers: 75002, 75003, 75004, 75005, 75006, 75007, 75008, 75009, 75010, 75011, 75012, 75014, 75015, 75016, 75017, 75018, 75019, 75020, 75021, 75022, 75023, 75024, 75025, 75026, 75030, 75032, 75035, 75036, 75037, 75040, 75042, 75044, 75047, 75048, 75049, 75051, 75053, 75060, 75064, 75100, 75102, 75104, 75105, 750111, 750114

Distribution Pattern:

Worldwide distribution to Belgium, China, Egypt, Germany, India, Korea, Republic of, Myanmar, Nigeria, Peru, Russian Federation, Rwanda, Singapore, Spain, Sweden, Taiwan, Thailand, and United Kingdom

Voluntary or Mandated:

Voluntary: Firm initiated