Philips Medical Systems (Cleveland) Inc: Medical Device Recall in 2016 - (Recall #: Z-2347-2016)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2016.

Data Source: FDA.
Product Description:

MX 16-slice SKD whole body computed tomography X-ray system. Imaging diagnostic tool.

Product Classification:

Class II

Date Initiated: April 20, 2016
Date Posted: August 17, 2016
Recall Number: Z-2347-2016
Event ID: 74402
Reason for Recall:

The following issues have been found in MX 16-slice systems with software version 1.1.4.21426: 1. During the filming operation on MX16-slice console software, the clipboard used for copying and pasting images is not cleared between patients. If the operator fails to copy the current patients images before pasting, a previous patients images may be present in the clipboard and be copied onto the film of the current patient. 2. During the Bolus Tracking scan, if the Auto Voice in Tracker scan is enabled, the Tracker scan will be aborted unexpectedly and the diagnostic scan after the tracker scan needs to be manually started. 3. If the scan protocol with SAS (Spiral Auto Start) function is selected to plan the scan, the SAS option may not be displayed on the Contrast tab of the scan protocol parameter area or the SAS option is displayed but not enabled as pre-configured in the protocol. This issue only occurs on the first helical scan after system startup that applies SAS function

Status: Terminated
Product Quantity: 80 units
Code Information:

Model No. 728132; S/N: EP16ER100003, EP16ER100004, EP16ER100005, EP16ER100007, EP16ER100008, EP16ER110003, EP16ER110004, EP16ER110005, EP16ER110006, EP16ER110007, EP16ER110008, EP16ER110009, EP16ER110010, EP16ER110011, EP16ER110012, EP16ER110013, EP16ER110014, EP16ER110015, EP16ER110016, EP16ER110017, EP16ER110018, EP16ER110019, EP16ER110020, EP16ER110021, EP16ER110022, EP16ER110023, EP16ER110024, EP16ER120001, EP16ER120002, EP16ER120003, EP16ER120004, EP16ER120005, EP16ER120006, EP16ER120007, EP16ER120008, EP16ER120010, EP16ER120012, EP16ER120015, EP16ER120016, EP16ER120017, EP16ER120018, EP16ER130001, EP16ER130002, EP16ER130005, EP16ER130006, EP16ER130007, EP16ER130008, EP16ER130009, EP16ER130010, EP16ER130011, EP16ER130012, EP16ER130013, EP16ER130014, EP16ER130015, EP16ER130016, EP16ER130018, EP16ER130025, EP16ER130026, EP16ER130029, EP16ER130033, EP16ER130034, EP16EB150009 & EP16EB150014. Model No.728135; S/N: EP16E120009, EP16E120011, EP16E120013, EP16ER130003, EP16ER130004, EP16ER130017, EP16ER130019, EP16ER130020, EP16ER130021, EP16ER130022, EP16ER130023, EP16ER130024, EP16ER130027, EP16ER130028, EP16ER130030, EP16ER130031 & EP16ER130032.

Distribution Pattern:

Products were sold in the following states: AL, AR, CA, FL, GA, IL, IN, MS, NC, ND, NJ, NV, NY, OH, PA, SC, TN & VA. Products were also shipped to the following countries: Albania, ALGERIA, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Bulgaria, Burundi, Cape Verde, Chile, China, Colombia, Congo, Democratic, Costa Rica, Cyprus, Dominican Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Hungary, India, INDONESIA, Ireland, Italy, Japan, Jordan, Latvia, Lebanon, Libyan Arab Jamahiri, Lithuania, Malawi, Mali, Mexico, Morocco, Myanmar, Namibia, Nepal, Netherlands, Nigeria, Oman, Palestine, Paraguay, Philippines, Poland, Romania, Russia, Senegal, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Uganda, United Kingdom, Venezuela, Viet Nam & Zambia.

Voluntary or Mandated:

Voluntary: Firm initiated