Philips Medical Systems (Cleveland) Inc: Medical Device Recall in 2016 - (Recall #: Z-2658-2016)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2016.

Data Source: FDA.
Product Description:

Ingenuity Core128, Model number 728323 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.

Product Classification:

Class II

Date Initiated: July 20, 2016
Date Posted: August 31, 2016
Recall Number: Z-2658-2016
Event ID: 74820
Reason for Recall:

After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at iso-center (0,0). If the region of interest is not visualized in the images, a CT rescan may be performed.

Status: Terminated
Product Quantity: 3
Code Information:

728323 -- Ingenuity Core128 320004 320053 320056

Distribution Pattern:

US Nationwide Distribution in the states of: AL,AR,AZ,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA, MD,ME,MI,MN,MO,MS,NC,ND,NE,NJ,NM, NV,NY,OH,OK,OR,PA,PR,SC,TN,TX,UT,VA,VT,WA,WI,WV,WY

Voluntary or Mandated:

Voluntary: Firm initiated