Philips Medical Systems (Cleveland) Inc: Medical Device Recall in 2018 - (Recall #: Z-0840-2018)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2018.

Data Source: FDA.
Product Description:

Brilliance Big Bore Oncology CTwith 4.2.0 software version. Model #' 728243 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories.

Product Classification:

Class II

Date Initiated: January 3, 2018
Date Posted: March 14, 2018
Recall Number: Z-0840-2018
Event ID: 79134
Reason for Recall:

Software issues including: Software errors that may result in Extended Field of View check message not displayed when Field of View is set >600 mm for a paused scan; Request for change to automatically turn on the center x/y 0 setting and the disable result rotation setting when turning on the CT simulation Exam Card; Software errors that may result in Tumor LOC allowing POI (Point of Interest) locking when a non-isocenter slice is displayed; Software errors that may result in partial set of images or inability to generate CT images.

Status: Terminated
Product Quantity: 11
Code Information:

Brilliance Big Bore Oncology CT, Model # 728243 Serial #s: 76025 76022 76024 76023 76016 76021 76008 76010 76002 76020 76011

Distribution Pattern:

Worldwide Distribution - US Nationwide

Voluntary or Mandated:

Voluntary: Firm initiated