Philips Medical Systems Gmbh, DMC: Medical Device Recall in 2020 - (Recall #: Z-1328-2020)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2020.

Data Source: FDA.
Product Description:

CombiDiagnost GCF, Software v. 1.0.5, Product Code 709030 - Product Usage: Xray system, diagnostic, fluoroscopic, general purpose.

Product Classification:

Class II

Date Initiated: December 19, 2019
Date Posted: February 26, 2020
Recall Number: Z-1328-2020
Event ID: 84773
Reason for Recall:

When using the Table Up/Down button, the system may experience Error 80, which locks the geometry in that specific state, requiring the intervention of a service Engineer. Additionally, the thermos switch, which handles power down of the unit in case of transformer overheating, was installed incorrectly at production.

Status: Ongoing
Product Quantity: 116
Code Information:

(domestic only) SN18061848 SN17071689 SN18051832 SN190251 SN17111759 SN17111752 SN18011793 10000020 SN17121783 SN17121785 SN18021816 SN18101890 SN19031950 SN17071688 SN18101903 SN17071683 SN17101722 SN17101729 SN17111758 SN17111763 SN17031622 SN17071692 SN17041642 SN18101886 SN18011794 SN19021934 SN19021943 SN17111770 SN18011802 SN19021942 SN17121779 SN17031618 SN17101733 SN17021613 SN17021613

Distribution Pattern:

Worldwide distribution  US Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, IN, MI, MN, MO, MT, NC, NJ, NY, OH, TX, UT, VA and countries of Canada, Australia, Austria, Bangladesh, Belgium, Bulgaria, Chile, China, Colombia, C¿te D'Ivoire, Cyprus, Czech Republic, Dominican Republic, Egypt, Ethiopia, France, Germany, Ghana, Hungary, Iran, Italy, Jordan, Kenya, Kuwait, Latvia, Lebanon, Libya, Luxembourg, Malaysia, Martinique, Mexico, Nepal, Netherlands, New Zealand, Nigeria, Oman, Philippines, Poland, Portugal, Romania, Russian Federation, Rwanda, Saudi Arabia, Serbia, South Africa, Spain, Switzerland, Tanzania, Thailand, Turkey, United Arab Emirates, United Kingdom, Viet Nam.

Voluntary or Mandated:

Voluntary: Firm initiated