Philips Medical Systems, Inc.: Medical Device Recall in 2015 - (Recall #: Z-1554-2015)
See the recall detail below. You can also see other recalls from the same firm in 2015.
Philips Healthcare DigitalDiagnost System X-Ray
Class II
The system is designed to emit a beep upon termination of an exposure. However, if the system has been powered on for over 12 hours, the system will no longer emit this signal. This is a failure to comply with 21CFR 1020.31(j).
Digital Diagnost Stationary X-ray System Release 4.0.0, 4.0.1, 4.0.2
Worldwide Distribution - Nationwide Distribution and the countries of Australia, Austria, Argentina, Canada, China, Chile, Colombia Czech Republic, Great Britain, Germany, Egypt, Finland, France, India, Ireland, Japan, Malaysia, Macedonia, Mexico, Netherlands, New Zealand, Nepal, Peru, Poland, Russia, Saudi Arabia Singapore, Switzerland, Sweden, Spain, United Arab Emirates and United Kingdom.
Voluntary: Firm initiated