PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: Medical Device Recall in 2026 - (Recall #: Z-1065-2026)
See the recall detail below. You can also see other recalls from the same firm in 2026.
Allura Xper FD10/10; Model Number: 722005;
Class II
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Model Number: 722005; System Product Name: Allura Xper FD10/10; UDI-DI: N/A; Serial Numbers: 93, 269, 530, 306, 307, 247, 53, 401, 141, 531, 184, 75, 276, 36;
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Honduras, India, Italy, Japan, Kenya, Korea, Republic of, Malaysia, Mexico, Nepal, Netherlands, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, Yemen.
Voluntary: Firm initiated