Philips Medical Systems North America Inc.: Medical Device Recall in 2014 - (Recall #: Z-1264-2014)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2014.

Data Source: FDA.
Product Description:

Philips Model EPIQ 7 Ultrasound System. Intended for diagnostic ultrasound imaging and fluid flow analysis. Catalogue numbers 795200 / 795201 / 795202 / 795204 / 795205 / 795206 EPIQ is intended for diagnostic ultrasound imaging and fluid flow analysis. The collection of Doppler velocity results values are used to assist in diagnosing the existence, location and severity of cardiac or vasculare stenosis.

Product Classification:

Class II

Date Initiated: February 28, 2014
Date Posted: March 26, 2014
Recall Number: Z-1264-2014
Event ID: 67685
Reason for Recall:

A problem has been detected with Philips Model EPIQ 5 and 7 Ultrasound System. The system will report incorrect Doppler velocities if you use High Pulse Repetition Frequency (HPRF) with multiple sample volume gates. While in High Pulse Repetition Frequency (HPRF) in Display Zoom Velocity Scale using manual spectral Doppler trace, the velocities in the velocity results table may be overstated b

Status: Terminated
Product Quantity: 1267 total :670 units in US and 1267 units Internationally (which includes 11 units in Canada)
Code Information:

All systems below software v1.1.2 Which includes software v1 and v 1.1 (All lot codes shipped prior to 2/28/2014).

Distribution Pattern:

Worldwide Distribution: US (nationwide) including states of: AL, AK, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NM, NJ, NM, NY, OH, OK, PA, SC, TN, TX, VA, WA, and WV; and countries of: Australia, Austria, Belgium, Canada, Chili, Ecuador, Egypt, Finland, France, Germany, Greece, India, Ireland, Italy, Japan, Korea Rep, Malaysia, Netherlands, New Zealand, Norway, Peru, Philippines, Portugal, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey and United Kingdom.

Voluntary or Mandated:

Voluntary: Firm initiated