Philips North America Llc: Medical Device Recall in 2021 - (Recall #: Z-1987-2021)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2021.

Data Source: FDA.
Product Description:

Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor 882442/882456 - Ingenuity TF PET/CT

Product Classification:

Class II

Date Initiated: May 1, 2021
Date Posted: July 7, 2021
Recall Number: Z-1987-2021
Event ID: 87974
Reason for Recall:

Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.

Status: Ongoing
Product Quantity: N/A
Code Information:

Model/ Serial Number/UDI: N/A System Code System Serial Number Unique Device Identifier (UDI#) 882442 2155 (01)00884838059443(21)2155 882442 2154 (01)00884838059443(21)2154 882442 2166 (01)00884838059443(21)2166 882442 2171 (01)00884838059443(21)2171 882442 2153 (01)00884838059443(21)2153

Distribution Pattern:

Worldwide distribution - US Nationwide and the countries of Canada, Mexico, AE, AR, AT, AU, BD, BE, BR, CH, CL, CN, CO, CU, CY, CZ, DE, DK, DZ, EE, EG, ES, FI, FR, GB, GE, GI, GL, GR, HK, HU, ID, IE, IL, IN, IQ, IR, IT, JO, JP, KE, KG, KH, KR, LA, LB, LK, LT, LY, MM, MN, MU, MV, MY, NI, NL, NO, NZ, OM, PA, PF, PH, PK, PL, PS, PT, QA, RE, RU, RW, SA, SE, SG , SI, SK, TH, TN, TR, UZ, VN, XK, YE, ZA, ZW.

Voluntary or Mandated:

Voluntary: Firm initiated