Ingenia Elition X-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body or extremities Model Number: 781358
Class II
Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury
S/N and UDI: US: 45245 (01)00884838088115(21)45245; 45247 (01)00884838088115(21)45247; 45261 (01)00884838088115(21)45261; 45263 (01)00884838088115(21)45263; 45264 (01)00884838088115(21)45264; 45269 (01)00884838088115(21)45269; OUS: 45215 (01)00884838088115; 45284 (01)00884838088115; 45253 (01)00884838088115; 45313 (01)00884838088115; 45265 (01)00884838088115; 45297 (01)00884838088115; 45218 (01)00884838088115; 45272 (01)00884838088115; 45281 (01)00884838088115; 45231 (01)00884838088115; 45278 (01)00884838088115; 45279 (01)00884838088115; 45245 (01)00884838088115; 45247 (01)00884838088115;
Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, MD, MD, NH, NY, OH, TN, TX, VA and the countries of Canada, Mexico, Australia, Austria, Chile, China, Czech Republic, Denmark, France, Germany, India, Indonesia, Italy, Japan, Korea, Republic of Lebanon, Netherlands, Romania, Russian Federation, Saudi Arabia, Spain, Turkey, Ukraine, United Kingdom.
Voluntary: Firm initiated