Philips North America Llc: Medical Device Recall in 2023 - (Recall #: Z-2654-2023)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2023.
Data Source: FDA.
Product Description:
EarlyVue VS30 Vitals Monitor, utilizing E01 (Oridion Microstream sidestream etCO2) and R21 (Masimo Rainbow RRa) simultaneously, Product Code 863380
Product Classification:
Class II
Date Initiated: August 31, 2023
Date Posted: October 11, 2023
Recall Number: Z-2654-2023
Event ID: 93020
Reason for Recall:
A software issue prevents the monitor from alarming when the high/low acoustic respiration rate (RRa) limits are violated.
Status: Ongoing
Product Quantity: 13 total devices
Code Information:
UDI-DI: 00884838091412; Serial Numbers: DE72401360 DE72401393 DE72400231 DE72400232 DE72400235 DE72400254 DE72400237 DE72400238 DE72400239 DE72409250 DE72409240 DE72409249
Distribution Pattern:
Worldwide - US Nationwide distribution in the states of MA, MI. The country of Brazil.
Voluntary or Mandated:
Voluntary: Firm initiated