Philips North America Llc: Medical Device Recall in 2024 - (Recall #: Z-1690-2024)
See the recall detail below. You can also see other recalls from the same firm in 2024.
Philips Patient Information Center (PIC) iX Uninterruptable Power Supply (UPS): 1. APC (Schneider Electric IT Corporation) 120V Uninterruptable Power Supply Model: SCL500RM1U 2. APC (Schneider Electric IT Corporation) 230V Uninterruptable Power Supply Model: SCL500RMI1U
Class II
Possible failure of Uninterruptable Power Supply (UPS) devices supporting the Patient Information Center (PIC) iX system, potential for the associated PIC iX system to shut down due to the lack of power. may lead to the delay in the detection of a change in a patient condition
1. SCL500RM1U (only serial number date codes prior to 2237*) 2. SCL500RMI1U (only serial number date codes prior to 2241*) Example unit serial number is 5S"2104"08943, meaning it was manufactured during week 04 of the year 2021. The serial number format is 5SYYWW123456.
Worldwide distribution - US Nationwide and the countries of Argentina, Aruba, Australia, Austria, Belgium, Bolivia, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Ecuador, Estonia, Fiji, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Ireland, Israel, Italy, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Malaysia, Mexico, Myanmar, Netherlands, New Zealand, Oman, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, Taiwan, United Kingdom, Utd. Ara Emir.
Voluntary: Firm initiated