Philips North America Llc: Medical Device Recall in 2024 - (Recall #: Z-1841-2024)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2024.
Data Source: FDA.
Product Description:
Achieva 3.0T TX for PET Model Number (REF): 781479
Product Classification:
Class II
Date Initiated: May 3, 2024
Date Posted: May 29, 2024
Recall Number: Z-1841-2024
Event ID: 94508
Reason for Recall:
The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire
Status: Ongoing
Product Quantity:
Code Information:
UDI-DI: N/A Serial Numbers: 43102
Distribution Pattern:
Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom.
Voluntary or Mandated:
Voluntary: Firm initiated