Philips North America Llc: Medical Device Recall in 2024 - (Recall #: Z-1859-2024)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2024.

Data Source: FDA.
Product Description:

MR RT Upgrades- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 781440

Product Classification:

Class II

Date Initiated: May 3, 2024
Date Posted: May 29, 2024
Recall Number: Z-1859-2024
Event ID: 94508
Reason for Recall:

The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire

Status: Ongoing
Product Quantity: 2 units
Code Information:

UDI-DI: N/A Serial Numbers: 70980, 71569

Distribution Pattern:

Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom.

Voluntary or Mandated:

Voluntary: Firm initiated