Philips North America Llc: Medical Device Recall in 2024 - (Recall #: Z-1861-2024)
See the recall detail below. You can also see other recalls from the same firm in 2024.
SmartPath to dStream for 3.0T Model Number (REF): 782145
Class II
The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire
UDI-DI: N/A Serial Numbers: 17377 34114 38294
Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom.
Voluntary: Firm initiated