Philips North America Llc: Medical Device Recall in 2024 - (Recall #: Z-3204-2024)
See the recall detail below. You can also see other recalls from the same firm in 2024.
SmartPath to dStream for XR and 3.0T- For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 1) 782113; 2) 781270; 3) 782129
Class II
Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.
Model UDI-DI: 782113 (01)00884838098909(21); 781270 (01)00884838095083(21); 782129 (01)00884838105805(21). Serial Numbers: 17271 17389 17439 17449 17504 24039 24077 34055 34072 34129 34262 38117 38166 38222
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Iceland, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Maldives, Mauritius, Mexico, Morocco, Netherlands, Pakistan, Peru, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Switzerland, Thailand, T¿rkiye, United Kingdom, Utd.Arab.Emir., Uzbekistan, Vietnam.
Voluntary: Firm initiated