Philips North America Llc: Medical Device Recall in 2025 - (Recall #: Z-2604-2025)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2025.

Data Source: FDA.
Product Description:

Philips CT systems labeled as the following: 1. CT 6000, Model Number: 728307.

Product Classification:

Class II

Date Initiated: August 13, 2025
Date Posted: October 1, 2025
Recall Number: Z-2604-2025
Event ID: 97468
Reason for Recall:

The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.

Status: Ongoing
Product Quantity: 6 units
Code Information:

1. Model Number: 728307; UDI-DI: 00884838104600; Serial Numbers: 87150, 87169, 87118, 87168, 87166, 87164.

Distribution Pattern:

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Mauritius, Mexico, Netherlands, Norway, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, T¿rkiye, United Kingdom, Utd.Arab.Emir., Vietnam.

Voluntary or Mandated:

Voluntary: Firm initiated