Philips North America: Medical Device Recall in 2023 - (Recall #: Z-0319-2024)
See the recall detail below. You can also see other recalls from the same firm in 2023.
Spectral CT on Rails, Software Version 5.1.0, Model Number 728334
Class II
This recall is part of a 2-issue recall for the same system. When using the IVC needle position button, on the IVC control box during a series of scans, the plan box may not move to the desired needle position. This issue could result in the gantry moving to a location the operator does not expect. If this occurs, it presents a risk that the scan could occur at an incorrect location, or that the unintended motion of the gantry could cause a collision with the user or needle. The other part of the recall involves a software crash when the CCT foot pedal is activated (as described in RES ID93363).
UDI-DI: 00884838103627; Lot Code: 1005
US Nationwide distribution in the state of MN.
Voluntary: Firm initiated