Philips North America: Medical Device Recall in 2024 - (Recall #: Z-0808-2024)
See the recall detail below. You can also see other recalls from the same firm in 2024.
Achieva 1.5T Conversion, Model No. 781283
Class III
An incorrect electrical shock warning label may have been placed on the Liquid Cooling Cabinet (LCC) of affected MR systems. The caution symbol on the incorrect label has a white background instead of yellow, and the label is missing the high voltage warning symbol. This presents a potential risk of electrical shock for service personnel.
Model No. 781283 UDI-DI 00884838009806 Serial Number 39016
Worldwide - US Nationwide distribution in the states of AR, CO, IL, MD, MS, NE, TX and the countries of Brazil, Canada, China, Germany, Greece, India, Mexico, and Saudi Arabia.
Voluntary: Firm initiated