Philips North America: Medical Device Recall in 2024 - (Recall #: Z-0973-2024)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2024.
Data Source: FDA.
Product Description:
Ingenia 1.5T (Product Number 781396) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system
Product Classification:
Class II
Date Initiated: January 6, 2022
Date Posted: February 14, 2024
Recall Number: Z-0973-2024
Event ID: 93737
Reason for Recall:
Magnetic materials were used for the wheels of trolley. Wheels with magnetic material might be attracted to MRI systems if the wheels are in the magnetic field.
Status: Ongoing
Product Quantity: 1
Code Information:
UDI-DI 00884838009820 Ingenia Serial number 41045 Accessory Serial Number 100018
Distribution Pattern:
Worldwide - US Nationwide distribution in the states of FL and NY. The countries of Denmark, Norway, and Sweden.
Voluntary or Mandated:
Voluntary: Firm initiated