Philips North America: Medical Device Recall in 2024 - (Recall #: Z-0975-2024)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2024.

Data Source: FDA.
Product Description:

Ingenia Ambition X (Product Number 781356) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system

Product Classification:

Class II

Date Initiated: January 6, 2022
Date Posted: February 14, 2024
Recall Number: Z-0975-2024
Event ID: 93737
Reason for Recall:

Magnetic materials were used for the wheels of trolley. Wheels with magnetic material might be attracted to MRI systems if the wheels are in the magnetic field.

Status: Ongoing
Product Quantity: 10
Code Information:

UDI-DI 00884838090040 Ingenia Serial number / Accessory Serial Number 47004 100020 47019 100024 47190 100022 47200 100025 47210 100019 47227 100063 47231 100056 47271 100032 47281 100028 47281 100031

Distribution Pattern:

Worldwide - US Nationwide distribution in the states of FL and NY. The countries of Denmark, Norway, and Sweden.

Voluntary or Mandated:

Voluntary: Firm initiated