Philips North America: Medical Device Recall in 2025 - (Recall #: Z-1930-2025)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2025.

Data Source: FDA.
Product Description:

Intera 1.5T Achieva Nova Product Number: 781172;

Product Classification:

Class II

Date Initiated: May 29, 2025
Date Posted: June 18, 2025
Recall Number: Z-1930-2025
Event ID: 96999
Reason for Recall:

identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a fire occurs, the risk to patients or operators may include inhalation of smoke, burns, and/or asphyxia which may lead to injury or even death. This issue could also lead to property damage.

Status: Ongoing
Product Quantity: 16 units (1 US, 15 OUS)
Code Information:

Product Number: 781172; UDI-DI: no UDI; Serial Numbers: 21218, 21217, 21216, 21289, 21349, 12146, 12063, 12086, 20069, 21027, 21198, 12099, 18641, 21476, 18623, 12078;

Distribution Pattern:

Worldwide - US Nationwide distribution in the states of AZ, CA, DE, MA, MI, MY, NE, NY, SD and the countries of Australia, Brazil, Ecuador, Germany, Iran, Italy, Japan, Lebanon, Libya, Mexico, Poland, Portugal, Spain.

Voluntary or Mandated:

Voluntary: Firm initiated