Philips Respironics, Inc.: Medical Device Recall in 2022 - (Recall #: Z-0494-2022)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2022.

Data Source: FDA.
Product Description:

Trilogy Evo Repair Kit, RP-Trilogy Evo Muffler Assembly, Material Number 1135257 The Trilogy Evo ventilator provides continuous or intermittent positive pressure ventilation for the care of individuals who require mechanical ventilation. Trilogy Evo is intended for pediatric through adult patients weighing at least 2.5 kg.

Product Classification:

Class I

Date Initiated: December 22, 2021
Date Posted: February 2, 2022
Recall Number: Z-0494-2022
Event ID: 89276
Reason for Recall:

Philips has identified a specific lot of non-conforming material manufactured by one of its suppliers. A Philips supplier incorrectly used polyester-based polyurethane (PE-PUR) sound abatement foam, a non-conforming material in the devices and repair kits. PE-PUR foam may degrade into particles which may enter the device s air pathway and be ingested or inhaled by the user. The foam degradation may be exacerbated by use of unapproved cleaning methods such as unapproved ozone cleaners, and the PE-PUR foam may off-gas certain chemicals during initial use of the device.

Status: Ongoing
Product Quantity: 205 kits
Code Information:

Lot Numbers 210414 - 210524

Distribution Pattern:

Worldwide Distribution: Distributed to the following US states: AR, CO, FL, IA, IL, KY, LA, MI, MT, OH, PA, TN, TX, UT, VA, WV, and WY; and Distributed to the following foreign countries: Argentina, Brazil, Romania, and Netherlands.

Voluntary or Mandated:

Voluntary: Firm initiated