Plasma Surgical Inc.: Medical Device Recall in 2012 - (Recall #: Z-1794-2012)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2012.
Data Source: FDA.
Product Description:
PlasmaJet Open Surgery Handpiece (Part of the PlasmaJet Neutral Plasma Surgery System) The PlasmaJet system is a neutral plasma surgery system that is designed for cutting and coagulation in open surgery and laparoscopic surgery.
Product Classification:
Class II
Date Initiated: April 27, 2012
Date Posted: June 20, 2012
Recall Number: Z-1794-2012
Event ID: 61951
Reason for Recall:
Potential failure in the glued seal between the stem and the tip of the handpiece leading to the possible leakage of a very small drop of distilled water coolant from the handpiece in use.
Status: Terminated
Product Quantity: 36 units
Code Information:
Lot Number 40041, exp. date February 2015 and Lot Number 40043, exp. date February 2015
Distribution Pattern:
Worldwide distribution: USA (nationwide) including states of: CA and NY; and country of: United Kingdom.
Voluntary or Mandated:
Voluntary: Firm initiated