PMBS, LLC: Medical Device Recall in 2016 - (Recall #: Z-1266-2016)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2016.

Data Source: FDA.
Product Description:

The S.C.O.R.E.S. Unit (self contained operating room equipment sterilization unit)

Product Classification:

Class II

Date Initiated: January 13, 2016
Date Posted: April 6, 2016
Recall Number: Z-1266-2016
Event ID: 73157
Reason for Recall:

Water retention issues. Units with filters that remain wet after the sterilization may compromised sterility.

Status: Terminated
Product Quantity: 58 units
Code Information:

ALL distributed S.C.O.R.E.S. units are affected.

Distribution Pattern:

FL, GA, MN, NY, OH, OR SC, and TN

Voluntary or Mandated:

Voluntary: Firm initiated