Polymer Technology Systems, Inc.: Medical Device Recall in 2019 - (Recall #: Z-1371-2019)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2019.

Data Source: FDA.
Product Description:

Lipid Panel test strips, REF 1710, also private labeled under Henry Schein as REF 570-0414

Product Classification:

Class II

Date Initiated: January 16, 2019
Date Posted: May 29, 2019
Recall Number: Z-1371-2019
Event ID: 82331
Reason for Recall:

Some lots of test strips do not fit tightly into the optical block, and may fall out during analysis, which could lead to an incorrect result being displayed.

Status: Terminated
Product Quantity: 67643 total
Code Information:

P832

Distribution Pattern:

Domestic distribution nationwide.

Voluntary or Mandated:

Voluntary: Firm initiated