Polymer Technology Systems, Inc.: Medical Device Recall in 2019 - (Recall #: Z-1372-2019)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2019.
Data Source: FDA.
Product Description:
Lipid Panel test strips (as a component of smart bundles), REF 2729, also private labeled under Henry Schein as REF 570-0412
Product Classification:
Class II
Date Initiated: January 16, 2019
Date Posted: May 29, 2019
Recall Number: Z-1372-2019
Event ID: 82331
Reason for Recall:
Some lots of test strips do not fit tightly into the optical block, and may fall out during analysis, which could lead to an incorrect result being di P832 Chemistry (Diagnostic) CHH 2 Lipid Panel test strips (as a component of smart bundles), REF 2729, also private labeled under Henry Schein as REF 570-0412
Status: Terminated
Product Quantity: 67643 total
Code Information:
Q802 Q811 Q813 Q816 Q820 Q823 Q831 Q903
Distribution Pattern:
Domestic distribution nationwide.
Voluntary or Mandated:
Voluntary: Firm initiated