Precision Medical, Inc.: Medical Device Recall in 2015 - (Recall #: Z-0272-2016)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2015.

Data Source: FDA.
Product Description:

Easy Dial Oxygen Regulator, 168708D and 168715D. Eight liter from .0.25 to 8 liters, 15 liter from 0.5 liter to 15 liters of oxygen per minute.

Product Classification:

Class II

Date Initiated: October 8, 2015
Date Posted: November 25, 2015
Recall Number: Z-0272-2016
Event ID: 72449
Reason for Recall:

These lots of Oxygen Dial Regulators could have the settings label off by one position. When the indicator arrow points to OFF position, the regulator is actually on the half liter setting (0.50), for the 15 liter. For the 8 liter, instead of being in the OFF position, the regulator will be in the quarter liter setting (0.25).

Status: Terminated
Product Quantity: 315
Code Information:

Lot numbers 072215 and 072315.

Distribution Pattern:

Distributed in the US to the states of PA, MA, MO, MT, ID, MI, MN, ND, KY, NY, NC, CA, AL, AZ, LA, OH, TX, GA, IN, and OK.

Voluntary or Mandated:

Voluntary: Firm initiated