Preservation Solutions, Inc.: Medical Device Recall in 2018 - (Recall #: Z-1387-2018)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2018.
Data Source: FDA.
Product Description:
Cold Storage Solution, 2L, labeled as Waters IGL Cold Storage Solution, Catalog No. CSS/2000/US; and, Belzer UW Cold Storage Solution, Catalog No. BUW_002. Intended for the hypothermic flushing and storage of specific organs in preparation for storage, transportation, and transplantation into a recipient.
Product Classification:
Class II
Date Initiated: February 26, 2018
Date Posted: April 25, 2018
Recall Number: Z-1387-2018
Event ID: 79378
Reason for Recall:
Bags of solution may have a small leak at the base of the injection port where the port is welded into the body of the bag..
Status: Terminated
Product Quantity: 31,104 units
Code Information:
UDI 00810805000163, and UDI 00810805000071.
Distribution Pattern:
Worldwide Distribution -- USA and Europe.
Voluntary or Mandated:
Voluntary: Firm initiated