ProSun International, LLC: Medical Device Recall in 2019 - (Recall #: Z-1176-2019)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2019.

Data Source: FDA.
Product Description:

ProSun Onyx 32 Sli Intensive (12 minutes) tanning bed, Model 32/1. Labeled as the following: PROSUN ONYX 32 SLI INTENSIVE RED, PROSUN ONYX 32 SLI INTENSIVE PEARL WHITE, ANYTIME FITNESS ONYX 32 SLI INTENSIVE BRONZE, PROSUN ONYX 32 SLI INTENSIVE BRONZE Tanning of the skin

Product Classification:

Class II

Date Initiated: December 10, 2018
Date Posted: April 24, 2019
Recall Number: Z-1176-2019
Event ID: 82346
Reason for Recall:

ProSun International LLC discovered discrepancies in their Quality Control Checklist documentation, which could result in high UV output.

Status: Terminated
Product Quantity: 177 devices of which 31 are affected
Code Information:

Serial Numbers: UNO32SLII4W0048, UNO32SLII1X0052, UNO32SLII1X0063, UNATF32SLII1X0028, UNO32SLII1X0064, UNO32SLII1X0065, UNO32SLII1X0067, UNO32SLII1X0068, UNO32SLII2X0071, UNO32SLII2X0072, UNO32SLII3V0070, UNO32SLII2X0075, UNO32SLII1X0069, UNO32SLII2X0073, UNATF32SLII2X0031, UNATF32SLII2X0030, UNO32SLII2X0084, UNO32SLII2X0082, UNO32SLII2X0083, UNO32SLII2X0081, UNO32SLII2X0091, UNO32SLII4W0050, UNO32SLII4W0047, UNO32SLII3W0011, UNO32SLII2X0079, UNO32SLII3X0095, UNO32SLII2X0094, UNO32SLII1X0056, UNO32SLII1X0057, UNATF32SLIIX0027, UNO32SLII3X0096

Distribution Pattern:

US Nationwide distribution to states of: CA, CO, FL, GA, IL, KY, LA, ME, MI, NC, NJ, NY, OH, OR, TN, TX, and WI.

Voluntary or Mandated:

Voluntary: Firm initiated