Pyng Medical Corporation: Medical Device Recall in 2015 - (Recall #: Z-0228-2016)
See the recall detail below. You can also see other recalls from the same firm in 2015.
MATResponder Tourniquet; Catalog number MATR: UPC 78353 00008; MATCombat Tourniquet, Catalog number MATC; UPC 78353 00010; MATResponder Tourniquet and MATCombat Tourniquet are essentially the same product with different color schemes manual nonpneumatic tourniquet devices. The device provides for emergency blood flow occlusion to the limbs
Class II
Once tested, it is required that the tourniquet be returned to its Ready position then placed in its packaging. Each tourniquet was tested for functionality, but in some cases, the tourniquet was not returned to its Ready position prior to packaging.
MATR: lot numbers; SL115802 (Exp 07/2024), SL12136 (Exp 08/2024), SL121360 (Exp 12/2024), SL114711 (Exp 01/2024) MATC - lot numbers : SL12131 (Exp 07/2024), SL121312 (Exp 08/2024), SL113171 (Exp 10/2023).
Worldwide Distribution - US Distribution and to the countries of : Colombia, France, Canada, Italy, Germany, Finland, United Kingdom, Australia, The Netherlands, Japan, Hungary, Indonesia, Israel, Mexico, Poland and Spain.
Voluntary: Firm initiated