Qualigen Inc: Medical Device Recall in 2024 - (Recall #: Z-0022-2025)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2024.
Data Source: FDA.
Product Description:
FastPack TSH Calibrator Kit, REF: 25000024, For use in conjunction with the FastPack IP TSH Immunoassay and FastPack IP System Analyzer
Product Classification:
Class II
Date Initiated: August 26, 2024
Date Posted: October 16, 2024
Recall Number: Z-0022-2025
Event ID: 95326
Reason for Recall:
Thyroid Stimulating Hormone (TSH) calibrator card has an incorrect barcode linked to incorrect information, that may allow successful system calibration, but the TSH results obtained could be higher than expected.
Status: Ongoing
Product Quantity: 60
Code Information:
UDI-DI: 00816467020099, Lot: 2405017-1, Expiration: 2025-05-15
Distribution Pattern:
US: TX, MS, NY, OH, MT, WA, PA, MN, SC, GA, NC, NV, WY, AK, SD, KS, NE, VT, AL, MA, OK, UT, WI, IA, WV. OUS: Germany, Switzerland
Voluntary or Mandated:
Voluntary: Firm initiated