Quidel Corporation: Medical Device Recall in 2021 - (Recall #: Z-2477-2021)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2021.

Data Source: FDA.
Product Description:

Lyra Direct SARS-CoV-2 Assay Emergency Use Authorization- Ref: M124. The Lyra Direct SARS-CoV-2 Assay is a real-time RT-PCR assay intended for the qualitative detection of nucleic acid from SARS-CoV-2 in nasal (NS), nasopharyngeal (NP), or oropharyngeal (OP) direct swab specimens from individuals suspected of COVID-19 by their healthcare provider

Product Classification:

Class II

Date Initiated: August 13, 2021
Date Posted: September 22, 2021
Recall Number: Z-2477-2021
Event ID: 88498
Reason for Recall:

Revised Instructions for Use: Potential for specimens from patients with high SARS-CoV-2 viral loads to be falsely interpreted as negative if target amplification occurs prior to a cycle-threshold (Ct) value of 5 when using the Thermo Fisher QuantStudio 7 Pro, the Applied Biosystems 7500 Fast Dx, the Applied Biosystems 7500, the Bio-Rad CFX96 Touch", the Roche LightCycler LC 480 II/ cobas z 480, or the Qiagen Rotor-Gene¿ MDx.

Status: Terminated
Product Quantity: 81,910 kits
Code Information:

All lots

Distribution Pattern:

Worldwide distribution - US Nationwide distribution including Puerto Rico and the countries of Canada, Panama, Virgin Islands, Germany, Honduras.

Voluntary or Mandated:

Voluntary: Firm initiated