R & D Systems, Inc.: Medical Device Recall in 2022 - (Recall #: Z-1390-2022)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2022.

Data Source: FDA.
Product Description:

Quantikine IVD Erythropoietin Human Serum Controls for use as quantitative controls for the determination of Erythropoietin concentrations in human serum and plasma. IFU, Part Number 750095

Product Classification:

Class III

Date Initiated: June 14, 2022
Date Posted: July 27, 2022
Recall Number: Z-1390-2022
Event ID: 90445
Reason for Recall:

The Instructions for Use (IFU, Part Number 750095) provided with the affected lots by the firm contains an error associated with product s CE marking. The CE mark contained in the IFU is followed by the Notified Body designation, British Standards Institution number 2797 (BSI-2797) for product inserts 750095.28 and 750095.29. The erroneous label CE2797 were in the IFU insert revision 750095.28 for the 408 affected lots that were shipped. No product was packed with erroneous insert revision 750095.29. Since the product is currently CE marked as a Class I - self-certified product under the EU In Vitro Diagnostics Directive, the Notified Body designation should not be specified. As part of the corrective action, the firm has notified the customers who received the products with erroneous labeling, and is conducting the replacements with the correct product insert containing the Insert Revision 750095.30

Status: Ongoing
Product Quantity: 400
Code Information:

Lot Code: UDI 00815762021671, Lot Codes: P301801; P304262; P306132; P310038; P314685; P320216; P322531 Expiration Date: 08/04/2023

Distribution Pattern:

Worldwide distribution - US Nationwide distribution in the states of FL, MN, UT, OH, MA, TX, PR, PA, NJ, CA, WI, LA, Puerto Rico, and the countries of CANADA, UNITED KINGDOM, ISRAEL, MEXICO, THAILAND, CHINA, HONG KONG, AUSTRALIA, INDONESIA.

Voluntary or Mandated:

Voluntary: Firm initiated