Radiometer America Inc: Medical Device Recall in 2013 - (Recall #: Z-0608-2013)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2013.
Data Source: FDA.
Product Description:
Qualicheck 1+, Aqueous Quality Control, Level 2, S7140, REF 944-042. The solution is used for quality check of blood gas analyzers.
Product Classification:
Class III
Date Initiated: June 6, 2006
Date Posted: January 2, 2013
Recall Number: Z-0608-2013
Event ID: 37530
Reason for Recall:
The barcode for the ABL77 on the insert incorrectly enters a control range for pH of 7.90-7.95 into the analyzer. The insert correctly states that the control range is 7.40-7.45.
Status: Terminated
Product Quantity: 840 Ampules
Code Information:
Lot R0040, expiration date February 2008; Lot R0042, expiration date April 2008.
Distribution Pattern:
Distributed to KY, NC, and NY.
Voluntary or Mandated:
Voluntary: Firm initiated