Radiometer Medical ApS: Medical Device Recall in 2020 - (Recall #: Z-1761-2020)
See the recall detail below. You can also see other recalls from the same firm in 2020.
ABL90 FLEX PLUS Analyzer, Model Number 393-092 - Product Usage: It is intended for use in a laboratory environment, near patient or point-of-care setting.
Class II
The firm received reports of occurrences where the barcode reader misinterpreted the contents of a locally printed barcode label used for entering patient ID or accession number into the analyzer in connection with a sample measurement. This could result is patient mixup and/or lost of sample, resulting in delayed medical treatment.
All software versions below 3.4 MR2
Worldwide distribution - US Nationwide distribution including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and the countries of Canada, Netherlands, China, Croatia, Czech Republic, Denmark, Germany, Spain, France, Hungary, India, Japan, South Korea, United Kingdom, Poland, Australia, South Africa, Singapore, Switzerland, Turkey, Norway, Austria, Italy, Sweden, Finland.
Voluntary: Firm initiated