Radiometer Medical ApS: Medical Device Recall in 2023 - (Recall #: Z-1394-2023)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2023.

Data Source: FDA.
Product Description:

AQURE REF 933-599 Software Versions 2.5.2 2.5.3 2.5.4 2.6.0 2.6.1 The AQURE system is intended to let you manage analytical devices and operator profiles. The user can associate patient data with test data. The system shows test results. The system receives data from connected devices at the point-of care or laboratory. It can send test results to the HIS/LIS. The system lets you send commands to selected devices. The system uses data related to the performance of devices, to tell users of issues to be managed. The AQURE system is intended for professional use.

Product Classification:

Class II

Date Initiated: March 8, 2023
Date Posted: April 26, 2023
Recall Number: Z-1394-2023
Event ID: 91917
Reason for Recall:

Due to potential software issue that may result in patient mix-up information.

Status: Ongoing
Product Quantity: 1,806 systems
Code Information:

Model Number: 933-599 UDI-DI Code: 05700699335999 Software Versions: 2.5.2 and above (Versions 2.5.2, 2.5.3, 2.5.4, 2.60, and 2.6.1)

Distribution Pattern:

Worldwide - U.S. Nationwide including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The countries of Argentina, Australia, Austria, Bahrain, Brazil, Chile, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Iraq, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Lithuania, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, Paraguay, Peru, Poland, Portugal, Qatar, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay.

Voluntary or Mandated:

Voluntary: Firm initiated