Randox Laboratories, Limited: Medical Device Recall in 2019 - (Recall #: Z-2084-2019)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2019.
Data Source: FDA.
Product Description:
RX Imola, Model Nos. RX4900
Product Classification:
Class II
Date Initiated: June 24, 2019
Date Posted: July 31, 2019
Recall Number: Z-2084-2019
Event ID: 83175
Reason for Recall:
There was a report of a malfunction with the sensor which resulted in an increase of temperature leading to melting of internal wires and scorch damage to the internal structure of the instrument.
Status: Terminated
Product Quantity: 27
Code Information:
All serial numbers
Distribution Pattern:
Domestic distribution to CA CT MA MD MO NC NJ NY OH PA UT WV and PR. Worldwide foreign distribution.
Voluntary or Mandated:
Voluntary: Firm initiated