Randox Laboratories Ltd.: Medical Device Recall in 2019 - (Recall #: Z-0840-2019)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2019.
Data Source: FDA.
Product Description:
Randox CALCIUM (Ca) COLORIMETRIC METHOD RX DAYTONA PLUS Cat. No. CA 8309 R1, Arsenazo Reagent 4 x 20 ml GTIN: 05055273208368
Product Classification:
Class II
Date Initiated: December 4, 2018
Date Posted: February 27, 2019
Recall Number: Z-0840-2019
Event ID: 81746
Reason for Recall:
Update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Calcium assay on RX instruments. Carryover to the Calcium reagent results may result in falsely elevated or decreased results to both patient and QC samples.
Status: Terminated
Product Quantity: 1 kit
Code Information:
All lot codes
Distribution Pattern:
WV, PR
Voluntary or Mandated:
Voluntary: Firm initiated