Randox Laboratories Ltd.: Medical Device Recall in 2021 - (Recall #: Z-1563-2021)
See the recall detail below. You can also see other recalls from the same firm in 2021.
Calibration Serum Level 3 (CAL 3) - multi-analyte mixture calibrator. Catalogue Number: CAL2351
Class II
Randox can confirm a positive bias for Total Bilirubin and Direct Bilirubin of up to +15% across the assay ranges when using calibrator CAL2351, lot 1162UE. The positive bias has been attributed to bilirubin instability in the calibrator and is evident in both Quality Control (QC) and Patient samples.
Batch / Lot number: 1162UE GTIN: 05055273200966
Domestic: IN, MA, CA, NJ, TX, NE, NY, VA, MD, MN, OR, IL, CT, DC, MS. Foreign: Albania, Argentina, Australia, Bangladesh, Bolivia, Bosnia & Herzegovina, Cameroon, Canada, Colombia, Cyprus, Denmark, Egypt, Estonia, Fiji, France, Germany, Greece, Hong Kong, Iraq, Italy, Ireland, Jordan, Kazakhstan, Kenya, Lithuania, Luxembourg, Mauritius, Mexico, Myanmar, Nepal, New Zealand, Pakistan, Palestine, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Thailand, The Netherlands, Turkey, UK, Ukraine, USA, and Vietnam.
Voluntary: Firm initiated