Randox Laboratories Ltd.: Medical Device Recall in 2024 - (Recall #: Z-1593-2024)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2024.
Data Source: FDA.
Product Description:
Microalbumin Calibrator Series (mALB CAL)-IVD used to calibrate the Microalbumin Assay Catalog Number: MA1567
Product Classification:
Class II
Date Initiated: January 31, 2024
Date Posted: April 24, 2024
Recall Number: Z-1593-2024
Event ID: 94138
Reason for Recall:
Microalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other methods and may misclassify patient results. Users may experience a shift in Quality Control and patient sample recovery of up to +12%.
Status: Ongoing
Product Quantity: 8 units
Code Information:
GTIN: 05055273204520 (1) Batch/Lot: 619282 / 1094MA-1098MA
Distribution Pattern:
US Nationwide distribution in the states of GA, MA, WV including PR.
Voluntary or Mandated:
Voluntary: Firm initiated