Raye's Inc.: Medical Device Recall in 2024 - (Recall #: Z-1595-2024)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2024.

Data Source: FDA.
Product Description:

SW Bari Lift & Transfer, Model Number 38060000

Product Classification:

Class II

Date Initiated: January 25, 2024
Date Posted: April 24, 2024
Recall Number: Z-1595-2024
Event ID: 93887
Reason for Recall:

Retrospective engineering analysis against current ISO standards for hoists has determined: 1. The device cannot physically perform at maximum documented weight specifications when a factor of safety of two times the maximum load is applied. 2. The device is not equipped with a safety device to ensure a person with disability would not fall in the event of a single-fault condition of the lifting device.

Status: Ongoing
Product Quantity: 1,149 units
Code Information:

UDI/DI 00845699001040, All serial numbers

Distribution Pattern:

Worldwide distribution - US Nationwide and the countries of Australia, UK.

Voluntary or Mandated:

Voluntary: Firm initiated