RAYSEARCH LABORATORIES AB: Medical Device Recall in 2015 - (Recall #: Z-1310-2015)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2015.

Data Source: FDA.
Product Description:

RayStation Radiation Therapy Treatment Planning System; -- RayStation 3.5, RayStation 4.0, RayStation 4.5 and RayStation 4.7. For RayStation 4.7, the issue applies also to machines with fixed jaws, regardless of MLC/jaw position. RayStation is a software system designed for treatment planning and analysis of radiation therapy.

Product Classification:

Class II

Date Initiated: March 4, 2015
Date Posted: April 1, 2015
Recall Number: Z-1310-2015
Event ID: 70686
Reason for Recall:

An issue with photon dose calculation for DMLC (Dynamic MLC) plans for machines where the MLC is positioned above the jaws, e.g. some Elekta linacs. The magnitude of the error depends on the beam model output factor corrections and on the individual DMLC plan characteristics.

Status: Terminated
Product Quantity: Domestic: 245 units
Code Information:

Build numbers 3.5.0.16, 3.5.1.6, 4.0.0.14, 4.0.1.4, 4.0.2.9, 4.0.3.4, 4.5.0.19, 4.5.1.14 and 4.7.0.15

Distribution Pattern:

Nationwide Distribution.

Voluntary or Mandated:

Voluntary: Firm initiated