RAYSEARCH LABORATORIES AB: Medical Device Recall in 2015 - (Recall #: Z-1679-2015)
See the recall detail below. You can also see other recalls from the same firm in 2015.
RaySearch RayStation 2.5, 3.0, 3.5 and 4.0; Radiation Therapy Treatment Product Usage: RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs. The intended users of RayStation shall be clinically qualified radiation therapy staff trained in using the system.
Class II
RaySearch Laboratories AB became aware of an issue with the dose calculation for some imported VMAT plans where controls points are defined with wide gantry angle spacing.
Software Version (Build numbers) 2.5.1.89, 3.0.0.251, 3.5.0.16, 3.5.1.6, 4.0.0.14, 4.0.1.4, 4.0.2.9, or 4.0.3.4
US Nationwide Distribution
Voluntary: Firm initiated