RAYSEARCH LABORATORIES AB: Medical Device Recall in 2015 - (Recall #: Z-2457-2015)
See the recall detail below. You can also see other recalls from the same firm in 2015.
RayStation 4.7, version 4.7.1, used with Structure Definition or Automatic Breast Planning. Radiation Therapy Treatment Planning System, for treatment planning and analysis of radiation therapy.
Class II
Issue with the algorithm for ROI contraction in RayStation 4.7 when non-uniform contraction distances are used. The ROI contraction tool uses six distances as input: right/left, inferior/superior and posterior/anterior. These contraction distances are, in error, pairwise reversed, i.e. right is interpreted as left, inferior is interpreted as superior, and posterior is interpreted as anterior.
software build numbers are 4.7.1.10
Distributed in CO, IL, MI, NC, OH,TN, and WA.
Voluntary: Firm initiated