RAYSEARCH LABORATORIES AB: Medical Device Recall in 2016 - (Recall #: Z-0721-2017)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2016.

Data Source: FDA.
Product Description:

RayStation 4.0, 4.5, 4.7 and 5.0; Radiation Therapy Treatment Planning System Product Usage: RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs. The intended users of RayStation shall be clinically qualified radiation therapy staff trained in using the system.

Product Classification:

Class II

Date Initiated: November 9, 2016
Date Posted: December 14, 2016
Recall Number: Z-0721-2017
Event ID: 75731
Reason for Recall:

Software anomaly; an issue was found with the proton Pencil Beam Scanning (PBS) dose calculation in RayStation 4.0, 4.5, 4.7 and 5.0. For treatment plans with a combination of range shifter, large air gap and beams that enter the patient surface at an oblique angle, the dose calculation accuracy may be less than expected.

Status: Terminated
Product Quantity: US: 66 units
Code Information:

Build numbers: 4.0.2.9, 4.0.3.4, 4.5.0.19, 4.5.1.14, 4.5.2.7, 4.7.0.15, 4.7.1.10, 4.7.2.5, 4.7.3.13, 4.7.4.4, 4.7.5.4, 5.0.0.37, 5.0.1.11 or 5.0.2.35

Distribution Pattern:

US Nationwide Distribution in the states of California, Florida, Illinois, Louisiana, Michigan, New Jersey, Tennessee, Texas, Washington

Voluntary or Mandated:

Voluntary: Firm initiated