RAYSEARCH LABORATORIES AB: Medical Device Recall in 2016 - (Recall #: Z-1712-2016)
See the recall detail below. You can also see other recalls from the same firm in 2016.
RayStation Therapy Treatment Planning System Stand-alone Software 3.0, 3.5, 4.0, 4.5 and 4.7., designed for treatment planning and analysis of radiation therapy.
Class II
A software issue with editing tools that use the left mouse button held down, for drawing in and interacting with the patient views and beams eye views. The views can become unsynchronized with the stored data if simultaneously right clicking, pressing Ctrl-S or Ctrl-Z while the left mouse button is held down. This bug does not affect dose computations, which are based on the stored system state.
3.0.0.251, 3.5.0.16, 3.5.1.6, 4.0.0.14, 4.0.1.4, 4.0.2.9, 4.0.3.4, 4.5.0.19, 4.5.1.14, 4.5.2.7, 4.7.0.15, 4.7.1.10, 4.7.2.5, 4.7.3.13 or 4.7.4.4
**Domestic** AR, AZ, CA, CT, CO, DE, FL, GA, IL, LA, MI, MD, MS, MA, NY, NV, NJ, OH, OR, PA, NC, RI, TN, WA & Washington D.C; **Internationally** Algeria, Australia, Austria, Belgium, Canada, China, France, French Polynesia, Germany, Iran, Italy, Japan, Netherlands, New Zealand, Norway, Republic of Korea, Spain, Sweden, Switzerland, Thailand & United Kingdom
Voluntary: Firm initiated